The seal peel tensile test determines the
strength of the seal at a specific place on the package which can be helpful
in setting the sealing parameters. This test involves measuring a segment
of the package along one of the seals and cutting the package so that
there is at least three inches (76mm)of material on each side of the heat
seal. This sample is then inserted into the grips of a tensile test instrument
with the heat seal centered between the two grips and the instrument initiated.
The results are reported as max. force, mean and standard deviation.
For further information on any of our services please contact us |
|||||
|
For all your Laboratory Testing
service requirements in any of the following areas contact Anecto Ltd:-
All our equipment (Qualmark, Thermotron, Votsch,
Lansmont, Derritron, Zwick) are fully calibrated and traceable to national
or international standards |
||||||